| Equipment / Parameter | Range | STD Equipment | |||
|---|---|---|---|---|---|
| I. Electrical- DC / LOW Frequency | DC Voltage Source | Up to 1.1 kV | Comparison to Multifunction Calibrator |
||
| DC Voltage Source | Fixed Values | 1 V 1.018 V 10 V |
Comparison to DC Voltage Stadard |
||
| DC Voltage Measure | 0 mV to 1 kV | Comparison to Multimeter | |||
| 1 kV to 10 kV | Comparison to High Voltage Meter with Probe | ||||
| DC Current Source | Up to 11 A | Comparison to Multifunction Calibrator |
|||
| 11 A to 1 000 A | Comparison to Multifunction Calibrator with Current Coil |
||||
| DC Current Measure | 0.1 nA to 1 A | Comparison to Multimeter | |||
| 1 A to 300 A | Comparison to Current Shunt Monitored with Multimeter | ||||
| Inductance Source | Fixed Points (1 kHz) | 100 uH to 10 H | Comparison to Standard Inductors | ||
| 1 kHz | 100 uH to 10 H | Comparison to Decade Inductors | |||
| Inductance Measure | 12 Hz to 1 kHz | 100 uH to 10 H | Comparison to RLC Digibridge | ||
| Resistance Source | Up to 330 MΩ | Comparison to Multifunction Calibrator |
|||
| 0 Ω to 100 GΩ | Comparison to Decade Resistance Box | ||||
| Resistance Source | Fixed Points | 0 Ω to 100 MΩ | Comparison to Multifunction Calibrator |
||
| 1 Ω to 10 MΩ | Comparison to Standard Resistors | ||||
| Resistance Measure | 0 Ω to 1 GΩ | Comparison to Multimeter | |||
| AC Resistance Measure | 12 Hz to 100 kHz | Up to 100 kΩ | Comparision to RLC Digibridge, Standard Resistors | ||
| AC Resistance Source | 10 Hz to 1 kHz | 1 Ω to 1 MΩ | Comparison to Standard Resistors | ||
| Capacitance Source | 1 kHz | 330 pF to 1.1 mF | Comparison to Multiproduct Calibrator | ||
| Capacitance Source | Fixed Points (1 kHz) | (1, 10, 100, 1 000) pF (10, 100, 1 000) nF |
Comparison to Standard Capacitor Set | ||
| Capacitance Measure | 1 kHz to 1 MHz | 1 pF to 1 nF | Comparison to Precision LCR Meter, RLC Digibridge | ||
| 1 kHz | 1 nF to 1 uF | ||||
| Electrical Stimulation of Thermocouple Indicators | Type B | 600 ℃ to 1 820 ℃ | Comparison to Multifunction Calibrator |
||
| Type C | 0 ℃ to 2 316 ℃ | ||||
| Type E | -250 ℃ to 1 000 ℃ | ||||
| Type J | -210 ℃ to 1 200 ℃ | ||||
| Type K | -200 ℃ to 1 372 ℃ | ||||
| Type L | -200 ℃ to 900 ℃ | ||||
| Type N | -200 ℃ to 1 300 ℃ | ||||
| Type R | 0 ℃ to 1 767 ℃ | ||||
| Type S | 0 ℃ to 1 767 ℃ | ||||
| Type T | -250 ℃ to 400 ℃ | ||||
| Type U | -200 ℃ to 600 ℃ | ||||
| Electrical Stimulation of RTD Indicators | Pt 385 | 100 Ω | -200 ℃ to 800 ℃ | Comparison to Multifunction Calibrator |
|
| Pt 3926 | 100 Ω | -200 ℃ to 630 ℃ | |||
| Pt 3916 | 100 Ω | -200 ℃ to 630 ℃ | |||
| Pt 385 | 200 Ω | -200 ℃ to 630 ℃ | |||
| Pt 385 | 500 Ω | -200 ℃ to 630 ℃ | |||
| Pt 385 | 1 kΩ | -200 ℃ to 630 ℃ | |||
| PtNi 385 | 100 Ω | -80 ℃ to 260 ℃ | |||
| Cu 427 | 10 Ω | -100 ℃ to 260 ℃ | |||
| AC Voltage Source | 10 Hz to 1 MHz | 220 uV to 220 V | Comparison to Multifunction Calibrator |
||
| 15 Hz to 1 kHz | 220 V to 1.1 kV | ||||
| AC Voltage Measure | 1 Hz to 300 kHz | Up to 10 mV | Comparison to Multifunction Calibrator |
||
| 1 Hz to 2 MHz | 10 mV to 10 V | ||||
| 1 Hz to 1 MHz | 10 V to 100 V | ||||
| 1 Hz to 100 kHz | 100 V to 700 V | ||||
| 60 Hz | 1 kV to 10 kV | Comparison to High Voltage Meter with Probe | |||
| AC Current Source | 10 Hz to 10 kHz | Up to 2.2 A | Comparison to Multifunction Calibrator |
||
| 10 Hz to 1 kHz | 2.2 to 11 A | ||||
| 60 Hz | 1 A to 550 A | Comparison to Multifunction Calibrator with Coil |
|||
| AC Current Measure | 10 Hz ~ 1 kHz | 5 uA to 100 uA | Comparison to Multimeter | ||
| 10 Hz ~ 5 kHz | 100 uA to 1 A | ||||
| 1 kHz | 1 A to 100 A | Comparison to Current Shunt with Multimeter | |||
| DC Power Source | 0.1 W to 11 kW | Comparison to Multifunction Calibrator |
|||
| AC Power Source | 60 Hz | 0.1 W to 12 kW | Comparison to Power Energy Calibrator | ||
| Oscilloscopes | Sine wave signal 50 Ω at 1 kHz 1 MΩ at 1 kHz |
1 mV to 6.6 V 1 mV to 130 V |
Comparison to Oscilloscope Calibrator, Active Head & RF Power Sensor | ||
| Pulse & Rise time 0.5 , 1 V p-p 1 Vp-p |
10 MHz 1 MHz |
||||
| Time Mark Output into 50 Ω | 2 ns to 5 s |
||||
| Leveled Sine Flatness | 50 kHz reference 50 kHz to 50 GHz |
||||
| Ⅱ. Electrical - RF/Microwave | Tuned RF Power, Relative Measure | 150 kHz to 26.5 GHz | -120 dB to 20 dB | Comparison to Spectrum Analyzer, Measuring Receiver | |
| RF Power Absolute Measure | 9 kHz to 50 GHz | -70 dBm to 20 dBm | Comparison to Measuring Receiver with Power Sensor | ||
| RF Power Source | 1 mHz to 20 MHz | -50 dBm to 20 dBm | Comparison to Signal Generator | ||
| 100 kHz to 50 GHz | -110 dBm to 10 dBm | ||||
| RF Power Meter Reference | 50 MHz | 1 mW | Comparison to Power Meter and Power Sensor | ||
| RF Power Sensor Calibration Factors | 9 kHz to 50 GHz | Referenced to 1 mW | |||
| Amplitude Modulation Depth Accuracy Measure | 150 kHz to 10 MHz (Rate: 50 Hz to 10 kHz) |
Depth: 5 % to 99 % |
Comparison to Spectrum Analyzer, Measuring Receiver | ||
| 10 MHz to 3 GHz (Rate: 90 Hz to 10 kHz) |
Depth: 20 % to 99 % |
||||
| 10 MHz to 3 GHz (Rate: 50 Hz to 100 kHz) |
Depth: 5 % to 20 % |
||||
| 3 GHz to 26.5 GHz (Rate: 50 Hz to 100 kHz) |
Depth: 20 % to 99 % | ||||
| Depth: 5 % to 20 % | |||||
| Amplitude Modulation Measure Flatness | 10 MHz to 26.5 GHz (Rate: 90 Hz to 10 kHz) |
Depth: 5 % to 99 % |
Comparison to PSA Series Spectrum Analyzer | ||
| FM Rejection (50 Hz to 3 kHz BW) |
250 kHz to 10 MHz (Rate: 400 Hz to 1 kHz) |
Depth: < 50 % Deviations: < 5 kHz |
|||
| 10 MHz to 26.5 GHz (Rate: 400 Hz to 1 kHz) |
Depth: < 50 % Deviations: < 50 kHz |
||||
| Residual AM | 250 kHz to 26.5 GHz | 50 Hz to 3 kHz BW | |||
| Harmonic Measurements | 30 Hz to 50 GHz | ≤ -80 dB to 0.1 dB | Comparison to Spectrum Analyzer | ||
| Harmonic Distortion | 20 Hz to 100 kHz | 0 dB to 100 dB | Comparison to Audio Analyzer | ||
| Frequency Modulation Deviation Accuracy Measure | 250 kHz to 10 MHz (Rate: 20 Hz to 10 kHz) |
Dev: ≤ 40 kHz |
Comparison to PSA Series Spectrum Analyzer, Measuring Receiver | ||
| 10 MHz to 26.5 GHz (Rate: 50 Hz to 200 kHz) |
Dev: ≤ 400 kHz |
||||
| Frequency Modulation Measure AM Rejection | 150 kHz to 26.5 GHz (50 Hz to 3 kHz BW) |
AM Depths: ≤ 50 % (Rate: 400 Hz or 1 kHz) |
Comparison to PSA Series Spectrum Analyzer | ||
| Residual FM | 100 kHz to 26.5 GHz | 50 Hz to 3 kHz BW | |||
| Phase Modulation Measure PM Deviation Accuracy |
100 kHz to 26.5 GHz | Deviations: > 0.7 rad Deviations: > 0.3 rad |
|||
| AM Rejection | 100 kHz to 26.5 GHz (50 Hz to 3 kHz BW) |
AM Depths: ≤ 50 % (Rate: 1 kHz) |
|||
| Residual PM | 100 kHz to 26.5 GHz | 50 Hz to 3 kHz BW | |||
| Amplitude Modulation Source | 10 kHz to 13.5 MHz | Rate : 50 Hz to 100 kHz Depths : (5 to 99)% | Comparison to AM/FM Test Source | ||
| Frequency Modulation Source | 10 kHz to 432 MHz | Rate : 20 Hz to 200 kHz Dev: ≤ 400 kHz peak |
|||
| Digital Modulation Source and Measure | EVM, Power Ramp Relative Accuracy | -0.11 dB ± 0.6 dB | Comparison to Signal Generator, Spectrum Analyzer | ||
| Phase and Frequency Error: | Phase Error | 0.01° to 180° | |||
| Peak Phase Error | -0.11 dB ± 0.6 dB | ||||
| Frequency Error | 0.001 Hz to 10 Hz | ||||
| Output RF Spectrum Relative RF Power Uncertainty | Due to Modulation Offsets ≤ 1.2 MHz Offsets ≥ 1.8 MHz |
-30 dB to 10 dB 1° to 15° 3° to 25° |
|||
| Due to Switching | -30 dB to 10 dB | 1° to 15° | |||
| Absolute RF Power Accuracy | 18° to 30° | ||||
| Excess Noise Ratio | 10 MHz to 26.5 GHz | ENR : 6 dB, 15 dB | Comparison to Noise Sources | ||
| S11 – Reflection Magnitude | 9 kHz to 50 GHz | Comparison to Network Analyzer, Calibration Kit and Verification Kit | |||
| S11 – Reflection Phase | |||||
| S21 – Transmission Magnitude | |||||
| S21 – Transmission Phase | |||||
| Ⅲ. Length and Dimensional metrology | Calipers | Up to 500 mm | Comparison to Gage Blocks | ||
| Height Gages | Up to 500 mm | Comparison to Gage Blocks, Surface Plate | |||
| Indicators | Up to 100 mm | Comparison to Gage Blocks | |||
| Micrometers | Up to 500 mm | ||||
| Ⅳ. Mass and Mass Related | Pressure | -15 to 10 000 psi | Comparison to Pressure Calibrator | ||
| Torque Tools | 0.01 N·m to 100 N·m | Comparison to Torque Calibrator | |||
| 60 lbf·ft to 600 lbf·ft | |||||
| Scale and Balances | 0 mg to 150 kg | Comparison to Standard Weights | |||
| Ⅴ. Photometry and Radiometry | Optical Wavelength Measure | 700 nm to 1 650 nm | Comparison to Wavelength Meter | ||
| Optical Power Measure | 850 nm | -60 dB to -6 dB | Comparison to Optical Test System | ||
| 1 310 nm | -110 dB to 10 dB | ||||
| 1 550 nm | |||||
| Ⅵ. Thermodynamic | Temperature Indicating Devices | -195 ℃ to 420 ℃ | Comparison to Standard Thermometer | ||
| Relative Humidity Indicating Devices | 10 % R.H. to 98 % R.H. | Comparison to Hygrometer | |||
| Ⅶ. Time and Frequency | Frequency Source | 0.1 Hz to 50 GHz | Comparison to GPS Receiver, Signal Generator | ||
| Frequency Measure | Fixed Point | 10 MHz | Comparison to GPS Receiver | ||
| Frequency Measure | 0.1 Hz to 46 GHz | Comparison to GPS, Frequency Receiver and Counter | |||
| Photo Tachometer | 0 to 100 000 RPM | Comparison to Waveform Generator, Multi-procduct Calibrator with LED | |||